15-7
Other Setup
Selecting OTHER SETUP option opens the following menu.
Figure 15-4 CO
2
Other Setup Menu
You can perform the following settings in this menu.
PUMP RATE Determines the sampling rate of the CO2 module pump.
N2O COMPEN N2O compensation
The presence of nitrous oxide causes the CO
2
value to be
higher than the actual value. Use this option to compensate
for the presence of N
2
O.
O2 COMPEN O2 compensation
The presence of a large concentration of oxygen causes the
CO
2
value to be lower than the actual value. Use this option
to compensate for the presence of O
2
.
Des COMPEN Desflurane compensation
The presence of desflurane causes the CO
2
value to be
higher than the actual value. Use this option to compensate
for the presence of desflurane.
NOTE
z The unit of N2O COMPEN, O2 COMPEN and Des COMPEN is “%”. The
concentration of N2O, O2 and Desflurane can be obtained from the anesthesia
machine.
z Set the N2O COMPEN, O2 COMPEN and Des COMPEN according to the
practical situation. Incorrect setting might incur errors from the actual value and
cause misdiagnosis.
BTPS Body Temperature and Pressure, Saturated
Options: ON and OFF.