10-27
10.6.6 Arr Threshold
Select ARR THRESHOLD >> in the ECG SETUP menu and the following menu appears. You
can change the threshold settings in this menu. An alarm will be triggered when a rate exceeds a
threshold.
Figure 10-18 ARR Threshold Setup
ARR Event Range Default Step Unit
ASYS DELAY 3 to 10 5 1 s
VTAC RATE 100 to 200 130 5 bpm
VTAC PVC 3 to 12 6 1 Beats
MULTIF. PVC 3 to 31 15 1 Beats
TACHY HI Adult: 100 to 300
Pediatric: 160 to 300
Adult: 100
Pediatric: 160
5 bpm
BRADY LO Adult: 15 to 60
Pediatric: 15 to 80
Adult: 60
Pediatric: 80
5 bpm
Vedere la pagina 130