14.4.2 IBP Pressure Zero Menu
You can select IBP PRESURE ZERO in IBP (1,2) SELECT menu to open the following menu.
Figure 14-9 IBP Pressure Zero Menu
Pressure Transducer Zeroing Procedure
1. Disconnect the transducer from the patient.
Figure 14-10 Pressure Transducer Zeroing
2. Adjust the 3-way stopcock to close the channel leading to the patient. The transducer is open
to the atmosphere through the stopcock.
3. Select CH1 ZERO or CH2 ZERO in IBP PRESSURE ZERO menu to start zeroing.
3-way
stopcock
Pressure
transducer
Pressure transducer
interface cable
Monitor