Circuit Principle
Pulse oximeter Service Manual 2-8
2.3 EMC Design
The following protections are used for EMC design in DPM1:
1. Every IC has one by-pass capacitor, some ICs have different rating capacitors to
enlarge the filter band;
2. Serial resistors on the communication lines connected with the PC, to reduce the
interference propagation;
3. RC low pass filter on most low frequency signals, to reduce high frequency
interference;
4. One ground plane, no discrimination between digital ground and analog ground;
5. Match resistor used on important signals, e.g. read/write, chip select etc., to
prevent oscillating or overshoot
Vedere la pagina 19