C-6
The sensor is
disconnected from
the patient or the
monitor, or the
sensor is not
properly connected.
Disconnect and
reconnect the sensor
as directed by the
instructions. If the
alarm remains, the
sensor or the cable
might have been
damaged.
SPO2 NO SENSOR
No Yes L
SpO
2
sensor is
reversed.
Disconnect and
reconnect the sensor
as directed by the
instructions. Pay
attention to the mark
on the sensor.
SPO2 WEAK
SIGNAL
No No L The pulse signals
detected by the
monitor are of poor
quality.
Move the sensor to a
site with better
signals.
WRONG SPO2
SENSOR
No No L The SpO
2
sensor is
incompatible to the
monitor, or is
damaged.
Stop using the sensor.
Vedere la pagina 101